Comparison of Portapres non-invasive blood pressure measurement in the finger with intra-aortic pressure measurement during incremental bicycle exercise
- 1 June 2002
- journal article
- clinical trial
- Published by Wolters Kluwer Health in Blood Pressure Monitoring
- Vol. 7 (3) , 179-183
- https://doi.org/10.1097/00126097-200206000-00006
Abstract
In order to evaluate the accuracy of continuous, non-invasive blood pressure measurements in the finger during stress, blood pressure was measured in six patients with a Portapres Model 2 during increasing levels of bicycle exercise, using simultaneously registered intra-aortic (aortic arch) pressure as a reference. Blood pressure was measured continuously and non-invasively in the left middle finger using the Portapres Model 2 device, at the same time invasively using a pigtail catheter situated in the aortic arch at rest and during and after standardized bicycle exercise. At the end of each stress level, in accordance with the protocol of the Association for the Advancement of Medical Intrumentation, systolic and diastolic blood pressure and pulse values were determined from the non-invasive (Portapres Model 2) and invasive measurements for 15 s each, and the mean values and standard deviations were calculated. Compared with simultaneous invasive blood pressure measurement in the aortic arch, the difference in measured blood pressure value (the value determined by the device minus the value determined invasively) at rest fell within the authorized normal range for German and European blood pressure measurement devices: systolic +1.7 ± 1.8 mmHg and diastolic –3.7 ± 6.2 mmHg. The difference in measured value increased with the stress level. Blood pressure and pulse can be determined reliably and accurately using the Portapres Model 2, non-invasively, both at rest and at low-to-moderate levels of stress.Keywords
This publication has 9 references indexed in Scilit:
- Verbesserung der koronaren Vasodilatationskapazität durch medikamentöse Lipidsenkung bei Patienten im Frühstadium der koronaren Atherosklerose mit eingeschränkter Koronarreserve und mäßiggradiger LDL-HypercholesterinämieClinical Research in Cardiology, 1998
- Fifteen years experience with finger arterial pressure monitoring: assessment of the technologyCardiovascular Research, 1998
- A software package for non-invasive, real-time beat-to-beat monitoring of stroke volume, blood pressure, total peripheral resistance and for assessment of autonomic functionAn updated and improved software version for Windows 95/NT and the complete biosignal electronics (ECG, ICG, beat-to-beat and oscillometric blood pressure and pulse oxymetry) will be supplied in a compact instrument by: CNSystems Medical Equipment Inc. Heinrichstrasse 22 A-8010 Graz, Austria, Europe. Tel: +43/316/3631-0; Fax: +43/316/3631-20; E-mail: sales@cnsystems.at; Internet: http.//www.cnsytems.atComputers in Biology and Medicine, 1998
- Aortic and Peripheral Blood Pressure During Isometric and Dynamic ExerciseInternational Journal of Sports Medicine, 1997
- COMPARISON OF INVASIVE AND NON-INVASIVE MEASUREMENT OF CONTINUOUS ARTERIAL PRESSURE USING THE FINAPRESBritish Journal of Anaesthesia, 1991
- Non‐invasive blood pressure monitoring during head‐up tilt using the Penaz principleActa Anaesthesiologica Scandinavica, 1990
- Comparison of Finapres non-invasive beat-to-beat finger blood pressure with intrabrachial artery pressure during and after bicycle ergometryJournal Of Hypertension, 1989
- Effects of peripheral vasoconstriction on the measurement of blood pressure in a fingerCardiovascular Research, 1985
- Direct Blood Pressure Measurement —Dynamic Response RequirementsAnesthesiology, 1981