Surface roughness and edge geometries in hemolysis with rotating disk flow
- 1 November 1981
- journal article
- research article
- Published by Wiley in Journal of Biomedical Materials Research
- Vol. 15 (6) , 923-939
- https://doi.org/10.1002/jbm.820150616
Abstract
Judging biocompatibility of materials with blood includes assessment of hemolysis resulting from flow in contact with those materials. Such hemolysis is influenced in part by the surface roughness and specific hydrodynamic features of the device used. Using a rotating‐disk device, with polyethylene disks and human blood, it was shown that roughness under 4 μm has negligible effect and that hemolysis increases sharply for roughness above about 11 μm. Gross roughness (65 μm) causes qualitatively different hemolysis kinetics but not as severe hemolysis as extrapolated from low‐roughness data. In the disk geometry, the corner is a key region which generates high hemolysis and thus minor alterations cause hemolytic variations which may tend to obscure materials influences. A series of corner‐beveled polycarbonate disks were tested in comparison with the normal square corner and were found always to cause greater hemolysis. Tapering the bevel inward, so the taper angle was on the order of 2°−4°, reduced hemolysis relative to the 45° bevel but still was more hemolytic than the flat disk. Evidence suggests that glassy plastics can be machined with more reproducibility at the corners than plastics above their Tg such as polyethylene. General device design problems regarding hemolysis are discussed.This publication has 14 references indexed in Scilit:
- Hemolytic Properties of Special Materials Exposed to a Shear Flow, and Plasma Changes with ShearBiomaterials, Medical Devices, and Artificial Organs, 1980
- Observations on Shear-Induced HemolysisBiomaterials, Medical Devices, and Artificial Organs, 1979
- Material Effects in Shear-Induced HemolysisBiomaterials, Medical Devices, and Artificial Organs, 1979
- Shear-Induced Hemolysis: Effects of Blood Chemistry (Including Aging in Storage) and Shearing SurfacesBiomaterials, Medical Devices, and Artificial Organs, 1976
- Hemolysis artifacts induced in rotating shear devicesJournal of Biomedical Materials Research, 1974
- Thrombogenicity of some biomedical materials: Platelet‐interface reactionsJournal of Biomedical Materials Research, 1969
- On the flow between a rotating and a stationary diskJournal of Fluid Mechanics, 1968
- The axially symmetric flow of a viscous fluid between two infinite rotating disksProceedings of the Royal Society of London. Series A. Mathematical and Physical Sciences, 1962
- De la nocivité de quelques matériaux employés dans la construction des organes artificiels et de leur nettoyage par des enzymes protéolytiquesVox Sanguinis, 1961
- HÆMOLYSIS CAUSED BY TUBING IN EXTRACORPOREAL CIRCULATIONThe Lancet, 1959