A multiple-capillary electrophoresis system for small-scale DNA sequencing and analysis
Open Access
- 15 December 1999
- journal article
- research article
- Published by Oxford University Press (OUP) in Nucleic Acids Research
- Vol. 27 (24) , 36e-36
- https://doi.org/10.1093/nar/27.24.e36
Abstract
A five-capillary system has been developed for DNA sequencing and analysis. The post-column fluorescence detector is based on a sheath-flow cuvette. The instrument provides uniform and continuous illumination of the samples. The cuvette virtually eliminates cross-talk in the fluorescence signal between capillaries. Discrete single-photon counting avalanche photodiodes provide high efficiency light detection. The instrument has detection limits (3σ) of 130 ± 30 fluorescein molecules injected onto each capillary. Over 650 bases of sequence at 98.8% accuracy were generated in 100 min at 50°C from M13mp18. Separation and detection of short tandem repeats proved efficient and accurate with the use of internal standards for direct comparison of migration times between capillaries.Keywords