Optimizing the precision in T1 relaxation estimation using limited flip angles
- 18 November 1987
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 5 (5) , 399-416
- https://doi.org/10.1002/mrm.1910050502
Abstract
In this article we describe the precision in the estimation of the spin–lattice relaxation time TI from MRI signals acquired for various flip angles with the repetition time TR held constant. We review the estimation procedure itself and present a model for the propagation of noise in the signal into the calculated TI. This model is verified by both Monte Carlo simulations and experimental data taken on image phantoms. Based on this model, we find that for a given TR/TI there exist two optimal flip angles that will minimize the uncertainty in the estimated TI. We also show how two optimal angles can be selected for a given range of TR/TI values. In addition, T1 estimation using the two optimal angles can be comparable to or better than using multiple evenly spaced angles. Finally, in an initial comparison with the two‐point saturation recovery method of calculating T1, results for equal total scanning time TRTot suggest that for T1 > 0.5 TRTot, the limited flip angle approach gives better TI precision whereas for TI < 0.5 TRTot, the saturation recovery approach is better. © Academic Press, Inc.This publication has 4 references indexed in Scilit:
- The precision of TR extrapolation in magnetic resonance image synthesisMedical Physics, 1986
- Strategies and tactics in NMR imaging relaxation time measurements. I. Minimizing relaxation time errors due to image noise—the ideal caseMagnetic Resonance in Medicine, 1985
- A new look at the method of variable nutation angle for the measurement of spin-lattice relaxation times using fourier transform NMRJournal of Magnetic Resonance (1969), 1977
- Application of Fourier Transform Spectroscopy to Magnetic ResonanceReview of Scientific Instruments, 1966