Reduction of Radiation Dose and Imaging Costs in Scoliosis Radiography
- 31 March 1988
- journal article
- research article
- Published by Wolters Kluwer Health in Spine
- Vol. 13 (4) , 409-412
- https://doi.org/10.1097/00007632-198804000-00007
Abstract
Photofluorography using a large-field image intensifier (Siemens Optilux 57) was applied to scoliosis radiography and compared with a full-size rare-earth screen/film technique. When scoliosis radiography (PA-projection) was performed on 25 adolescent patients, the photofluorographs were found to be of comparable diagnostic quality with full-size films. A close correspondence between the imaging techniques was found in the Cobb angle measurements as well as in the grading of rotation with the pedicle method. The use of photofluorography results in a radiation dose reduction of about one-half and considerable savings in direct imaging costs and archive space. In our opinion the method is particularly well-suited for follow-up and screening evaluation of scoliosis, but in tall patients the image field size of 40 X 40 cm restricts its usefulness as initial examination.Keywords