Clinical performance of the Reichert AT550: a new non‐contact tonometer
- 24 November 2002
- journal article
- clinical trial
- Published by Wiley in Ophthalmic and Physiological Optics
- Vol. 22 (6) , 560-564
- https://doi.org/10.1046/j.1475-1313.2002.00077.x
Abstract
The aim of the present study was to assess the level of accuracy for measurements of intra-ocular pressure (IOP) obtained with a new non-contact tonometer (NCT) the Reichert AT550. Measurements were compared against those obtained with the Reichert Xpert Plus, Goldmann applanation tonometer and Perkins tonometer. Thirty-five university students were assessed with the four tonometers in a randomised order, with non-contact tonometry performed first. Each of the four measurement devices had its own trained clinical observer. Plots of differences of IOP as a function of the mean for each pair of instruments were obtained. No statistically significant differences were found when comparing the AT550 NCT with contact applanation tonometry (AT) (p> 0.05), displaying the closest level of agreement (as represented by the lowest mean difference and the narrowest confidence interval) with the Goldmann tonometer (limits of agreement, 0.12+/-2.17). In conclusion, readings of IOP with the AT550 NCT are clinically comparable with those obtained with Goldmann tonometry in a population with IOP within the normal range.Keywords
This publication has 17 references indexed in Scilit:
- The Pulsair 3000 tonometer — how many readings need to be taken to ensure accuracy of the average?Ophthalmic and Physiological Optics, 2001
- Twenty-four-hour time course of intraocular pressure in healthy and glaucomatous africans: relation to sleep patternsOphthalmology, 2001
- Intraocular Pressure and Photorefractive KeratectomyCornea, 2001
- A clinical comparison of the Xpert non-contact tonometer with the Goldmann applanation tonometer after penetrating keratoplastyActa Ophthalmologica Scandinavica, 2000
- Reliability of intraocular pressure measurements after myopic excimer photorefractive keratectomyOphthalmology, 1998
- Intraocular pressure asymmetry is not a clinically-significant feature when using the PULSAIR non-contact tonometerOphthalmic and Physiological Optics, 1997
- The diurnal variation of intraocular pressure in non-glaucomatous subjects: relevance in a clinical contextOphthalmic and Physiological Optics, 1997
- Comparison of the Performance of the Nidek NT-2000 Noncontact Tonometer with the Keeler Pulsair 2000 and the Goldmann Applanation TonometerOptometry and Vision Science, 1997
- Comparing methods of measurement: why plotting difference against standard method is misleadingThe Lancet, 1995
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986