Quantification of the magnification and distortion effects of a pediatric flexible video-bronchoscope
Open Access
- 10 February 2005
- journal article
- research article
- Published by Springer Nature in Respiratory Research
- Vol. 6 (1) , 16
- https://doi.org/10.1186/1465-9921-6-16
Abstract
Background: Flexible video bronchoscopes, in particular the Olympus BF Type 3C160, are commonly used in pediatric respiratory medicine. There is no data on the magnification and distortion effects of these bronchoscopes yet important clinical decisions are made from the images. The aim of this study was to systematically describe the magnification and distortion of flexible bronchoscope images taken at various distances from the object. Methods: Using images of known objects and processing these by digital video and computer programs both magnification and distortion scales were derived. Results: Magnification changes as a linear function between 100 mm (×1) and 10 mm (×9.55) and then as an exponential function between 10 mm and 3 mm (×40) from the object. Magnification depends on the axis of orientation of the object to the optic axis or geometrical axis of the bronchoscope. Magnification also varies across the field of view with the central magnification being 39% greater than at the periphery of the field of view at 15 mm from the object. However, in the paediatric situation the diameter of the orifices is usually less than 10 mm and thus this limits the exposure to these peripheral limits of magnification reduction. Intraclass correlations for measurements and repeatability studies between instruments are very high, r = 0.96. Distortion occurs as both barrel and geometric types but both types are heterogeneous across the field of view. Distortion of geometric type ranges up to 30% at 3 mm from the object but may be as low as 5% depending on the position of the object in relation to the optic axis. Conclusion: We conclude that the optimal working distance range is between 40 and 10 mm from the object. However the clinician should be cognisant of both variations in magnification and distortion in clinical judgements.Keywords
This publication has 18 references indexed in Scilit:
- Quantitative Assessment of Tracheal Collapsibility in Infants with TracheomalaciaAmerican Journal of Respiratory and Critical Care Medicine, 2004
- Complications of flexible bronchoscopy in children: prospective study of 1,328 proceduresEuropean Respiratory Journal, 2002
- Airway obstruction in children with congenital heart disease: Assessment by flexible bronchoscopyPediatric Pulmonology, 2002
- Series of laryngomalacia, tracheomalacia, and bronchomalacia disorders and their associations with other conditions in childrenPediatric Pulmonology, 2002
- A new method to perform quantitative measurement of bronchoscopic images.European Respiratory Journal, 2001
- Measurement of Lesions by Endoscopy: An OverviewEndoscopy, 1995
- Endoscopic measurement of lesion size: Improved accuracy with image processingPublished by Elsevier ,1994
- Left mainstem bronchopexy for severe bronchomalaciaJournal of Pediatric Surgery, 1991
- Intraluminal Airway Inflammation in Chronic Bronchitis: Characterization and Correlation with Clinical ParametersAmerican Review of Respiratory Disease, 1989
- Bronchomalacia in ChildrenClinical Pediatrics, 1970