Computer modeling of visual impairment caused by intraocular lens misalignment
- 1 January 1999
- journal article
- Published by Wolters Kluwer Health in Journal of Cataract & Refractive Surgery
- Vol. 25 (1) , 100-105
- https://doi.org/10.1016/s0886-3350(99)80019-4
Abstract
Purpose: To evaluate a computer program to modulate the visual impairment caused by intraocular lens (IOL) misalignment and visualize results obtained by numerical calculations. Setting: Department of Ophthalmology, Medical Faculty of Charles University, Prague, Czech Republic. Methods: The optic imagery of a Landolt circle was calculated using a ray-tracingcomputer program. Visual aberrations resulting from a decentered and/or tilted IOL were studied using this program and compared with theoretical calculations. Results: The IOL decentration and/or tilt shifted the postoperative refractive errorstoward myopia and astigmatism (oblique). The combination of IOL decentration and tilt produced a refractive error that depended on the relationship between the geometrical axes of the decentration and tilt. The refractive error can be enhanced or diminished depending on the relationship of these axes. Conclusions: These findings verify the results calculated by paraxial vergence equations. A ray-tracing program simulated the optic imagery for various kinds of IOL misalignment and IOL optic properties.Keywords
This publication has 8 references indexed in Scilit:
- Tilt and decentration of the implanted posterior chamber intraocular lensJournal of Cataract & Refractive Surgery, 1991
- Rapid Determination of Intraocular Lens Tilt and Decentration Through the Undilated PupilOphthalmology, 1990
- Effects of intraocular lens position errors on postoperative refractive errorJournal of Cataract & Refractive Surgery, 1990
- In Vivo Measurement of Posterior Chamber Intraocular Lens Decentration and TiltArchives of Ophthalmology (1950), 1990
- Optical Design of Intraocular Lenses. III. On-Axis Performance in the Presence of Lens DisplacementOptometry and Vision Science, 1989
- Measurement of postoperative intraocular lens tilting and decentration using Scheimpflug imagesJournal of Cataract & Refractive Surgery, 1989
- Refractive Errors Induced by Displacement of Intraocular Lenses within the Pseudophakic EyeOptometry and Vision Science, 1989
- Measurement of intraocular lens decentration and tilt in vivoJournal of Cataract & Refractive Surgery, 1988