Patient-ventilator asynchrony
- 1 February 2001
- journal article
- review article
- Published by Wolters Kluwer Health in Current Opinion in Critical Care
- Vol. 7 (1) , 28-33
- https://doi.org/10.1097/00075198-200102000-00005
Abstract
The basic mechanism of patient-ventilator asynchrony is the mismatching between neural inspiratory and mechanical inspiratory time. Alterations in respiratory drive, timing, respiratory muscle pressure, and respiratory system mechanics influence the interaction between the patient and the ventilator. None of the currently available partial ventilatory support modes are exempt from problems with patient-ventilator asynchrony. Ventilator triggering design in the trigger phase and the set variables in the post-trigger phase contribute to patient-ventilator interaction. The set inspiratory flow rate in the post-trigger phase for assist-control volume cycled ventilation affects patient-ventilator asynchrony. Likewise, the initial pressure rise time, the pressure support level, and the flow-threshold for cycling off inspiration for pressure support ventilation are important factors affecting patient-ventilator asynchrony. Current investigations have advanced our understanding in this area; however, its prevalence and the extent to which patient-ventilator asynchrony affect the duration of mechanical ventilation remain unclear.Keywords
This publication has 20 references indexed in Scilit:
- Analysis of the mechanisms of expiratory asynchrony in pressure support ventilation: a mathematical approachJournal of Applied Physiology, 2000
- Response of Ventilator-dependent Patients to Different Levels of Pressure Support and Proportional AssistAmerican Journal of Respiratory and Critical Care Medicine, 1999
- Effect of Ventilator Flow Rate on Respiratory Timing in Normal HumansAmerican Journal of Respiratory and Critical Care Medicine, 1999
- Cycling of Inspiratory and Expiratory Muscle Groups with the Ventilator in Airflow LimitationAmerican Journal of Respiratory and Critical Care Medicine, 1998
- Patient-Ventilator Trigger Asynchrony in Prolonged Mechanical VentilationChest, 1997
- Patient-ventilator flow dyssynchronyCritical Care Medicine, 1997
- Comparison of assisted ventilator modes on triggering, patient effort, and dyspnea.American Journal of Respiratory and Critical Care Medicine, 1997
- Partitioning of inspiratory muscle workload and pressure assistance in ventilator-dependent COPD patients.American Journal of Respiratory and Critical Care Medicine, 1996
- Variability of patient-ventilator interaction with pressure support ventilation in patients with chronic obstructive pulmonary disease.American Journal of Respiratory and Critical Care Medicine, 1995
- On the regulation of depth and rate of breathingThe Journal of Physiology, 1972