Real-Time Data Rebinning in PET to Obtain Uniformly Sampled Projections
- 1 January 1985
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Nuclear Science
- Vol. 32 (1) , 811-817
- https://doi.org/10.1109/tns.1985.4336946
Abstract
In many PET systems, the detector arrays can move during data acquisition in order to augment the sampling rate. This increases resolution and eliminates aliasing noise. With discrete detector movement there is a fairly long minimum scan duration. Continuous precession (wobble) allows shorter scans but increases in on-line memory needed. We modified the Positome III at the Montreal Neurological Institute to allow it to precess continuously, and interpolate the coincidence data in real time into uniformly sampled parallel projections. We are now able to perform 12 mm resolution scans of any frame rate greater than 2 seconds using 30% of the memory previously needed. Furthermore, we have reduced the reconstruction time by one third.Keywords
This publication has 11 references indexed in Scilit:
- Evaluation of the Performance Characteristics of the PC 4600 Positron Emission TomographJournal of Computer Assisted Tomography, 1984
- Performance Study of the PC-384 Positron Camera System for Emission Tomography of the BrainJournal of Computer Assisted Tomography, 1984
- Performance Figures and Images from the Therascan 3128 Positron Emission TomographIEEE Transactions on Nuclear Science, 1984
- PETT VIJournal of Computer Assisted Tomography, 1982
- A Fractional Address Accumulator for Fast Back-ProjectionIEEE Transactions on Nuclear Science, 1981
- Algorithms for Fast Back- and Re-Projection in Computed TomographyIEEE Transactions on Nuclear Science, 1981
- Effects of Wobbling Motion on Image Quality in Positron TomographyIEEE Transactions on Nuclear Science, 1981
- Quantitation in Positron Emission Computed TomographyJournal of Computer Assisted Tomography, 1980
- Data Collection for Cross-Sectional Image Reconstruction by a Moving Ring of Positron Annihilation DetectorsJournal of Computer Assisted Tomography, 1979
- Sampling Requirements and Detector Motion for Positron Emission TomographyIEEE Transactions on Nuclear Science, 1979