Quantitating fluorescence intensity from fluorophores: Practical use of MESF values
- 1 July 2002
- journal article
- Published by National Institute of Standards and Technology (NIST) in Journal of Research of the National Institute of Standards and Technology
- Vol. 107 (4) , 339-53
- https://doi.org/10.6028/jres.107.027
Abstract
The present work uses fluorescein as the model fluorophore and points out critical steps in the use of MESF (Molecules of Equivalent Soluble Fluorophores) values for quantitative flow cytometric measurements. It has been found that emission spectrum matching between a reference solution and an analyte and normalization by the corresponding extinction coefficient are required for quantifying fluorescence signals using flow cytometers. Because of the use of fluorescein, the pH value of the medium is also critical for accurate MESF assignments. Given that the emission spectrum shapes of microbead suspensions and stained biological cells are not significantly different, the percentage of error due to spectrum mismatch is estimated. We have also found that the emission spectrum of a microbead with a seven-methylene linker between the fluorescein and the bead surface (bead7) provides the best match with the spectra from biological cells. Therefore, bead7 is potentially a better calibration standard for flow cytometers than the existing one that is commercially available and used in the present study.Keywords
This publication has 11 references indexed in Scilit:
- Quantitating fluorescence intensity from fluorophore: The definition of MESF assignmentJournal of Research of the National Institute of Standards and Technology, 2002
- Terminology and nomenclature for standardization in quantitative fluorescence cytometryCytometry, 1998
- Quantitation of CD38 activation antigen expression on CD8+ T cells in HIV-1 infection using CD4 expression on CD4+ T lymphocytes as a biological calibratorCytometry, 1998
- Quantitation of CD38 expression using QuantiBRITE? beadsCytometry, 1998
- Spectral properties of the prototropic forms of fluorescein in aqueous solutionJournal of Fluorescence, 1996
- Standardizing flow cytometry: construction of a standardized fluorescence calibration plot using matching spectral calibratorsCytometry, 1996
- Experimental tests of a simple diffusion model for the estimation of scattering and absorption coefficients of turbid media from time-resolved diffuse reflectance measurementsApplied Optics, 1992
- Interlaboratory study of cellular fluorescence intensity measurements with fluorescein‐labeled microbead standardsCytometry, 1991
- Analysis of T cell subsets in normal adultsJournal of Immunological Methods, 1987
- Right angle light scatter: a necessary parameter in flow cytofluorimetric analysis of human peripheral blood mononuclear cellsJournal of Immunological Methods, 1983