Influence of the suture diameter on the tensile strength of polypropylene monofilaments and its relationship to the USP specification
- 1 September 1984
- journal article
- research article
- Published by Wiley in Journal of Biomedical Materials Research
- Vol. 18 (7) , 781-787
- https://doi.org/10.1002/jbm.820180708
Abstract
Ethylene gas sterilized polypropylene monofilaments were tested in a straight pull as well as in a conventional knot pull tensile test. In either case an analysis of the results indicates that the occurrence of “surface steps” may explain the relationship between strength and suture diameter. The present investigation not only shows that the polypropylene specification could be established on a more fundamental basis (using 60% of the ultimate tensile strength as the generalized criterion) but also suggests that other sutures should be evaluated similarly.This publication has 7 references indexed in Scilit:
- The solid-state polymerization and physical properties of bis(ethyl urethane) of 2,4-hexadiyne-1,6-diol: 3. Mechanical propertiesPolymer, 1983
- Creep rupture behavior of polypropylene suture material and its applications as a time-release mechanismJournal of Biomedical Materials Research, 1983
- Strengthening of lamellar vs. equiaxed Ag-Cu eutecticActa Metallurgica, 1970
- Some problems of the radiation chemistry of polypropylenePolymer Science U.S.S.R., 1968
- The structure and properties of drawn pearliteActa Metallurgica, 1966
- The plastic deformation of polycrystalline aggregatesPhilosophical Magazine, 1962
- The Deformation and Ageing of Mild Steel: III Discussion of ResultsProceedings of the Physical Society. Section B, 1951