The START Trial - measurements in semi-anatomical breast and chest wall phantoms
- 21 June 2001
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 46 (7) , 1937-1948
- https://doi.org/10.1088/0031-9155/46/7/314
Abstract
This paper describes dosimetry measurements performed prior to departments entering patients into the START Trial, a breast fractionation trial. Absolute and relative doses were measured in semi-anatomical breast and chest wall phantoms, as part of a quality assurance programme visit. Doses were measured using an ionization chamber and the resulting distributions were compared with those calculated by the department. The mean ratio of measured to calculated dose at the START reference point was found to be 0.981 for the breast phantom and 0.978 for the chest wall phantom. This average measured dose was significantly less than the prescribed dose (p < 0.001). Differences were found between 2D and 3D planning systems and for departments using cobalt 60 beams. A number of departments had deviations of greater than 4%, which was the tolerance applied for this trial. It is essential for dose measurements of this type to be performed for randomized clinical trials involving radiotherapy, particularly where dose fractionation regimes are being compared.Keywords
This publication has 7 references indexed in Scilit:
- Breast radiotherapy phantom design for the START trial. START trial management group.The British Journal of Radiology, 2000
- Three-dimensional dose distribution of tangential breast irradiation: results of a multicentre phantom dosimetry studyRadiotherapy and Oncology, 2000
- Tangential breast irradiation: a multi-centric intercomparison of dose using a mailed phantom and thermoluminescent dosimetry.Radiotherapy and Oncology, 1999
- Three-dimensional dose distribution of tangential breast treatment: a national dosimetry intercomparisonRadiotherapy and Oncology, 1991
- Code of Practice for high-energy photon therapy dosimetry based on the NPL absorbed dose calibration servicePhysics in Medicine & Biology, 1990
- Superficial doses in treatment of breast with tangential fields using 4 MV x-raysInternational Journal of Radiation Oncology*Biology*Physics, 1977
- Depth dose in skin for obliquely incident60Co radiationThe British Journal of Radiology, 1972