Blood gas analyses on equine blood: required correction factors
- 1 November 1991
- journal article
- Published by Wiley in Equine Veterinary Journal
- Vol. 23 (6) , 410-412
- https://doi.org/10.1111/j.2042-3306.1991.tb03752.x
Abstract
Correction factors have been determined to obtain the best estimates of PO2, PCO2 and pH in equine blood with standard blood gas and pH electrodes. There was a significant difference between the PO2 readings for tonometred blood of most horses and the equilibrating gas. Thus, if the PO2 electrode is calibrated with a gas, an electrode correction factor should be obtained by tonometring a blood sample from each horse. This factor was not dependent on packed cell volume. No such correction is required for the PCO2 electrode. If the animal's temperature differs from that of the analyser, the PO2, PCO2 and pH values must be corrected to the animal's body temperature. Temperature correction factors determined for equine blood were similar to those for human blood. Failure to make temperature corrections can result in errors for PO2 and PCO2 of 6 to 7 per cent per degree of temperature difference.Keywords
This publication has 9 references indexed in Scilit:
- Blood gas measurements during exercise: errors due to temperature correctionJournal of Applied Physiology, 1989
- The effect of temperature on the oxygen equilibrium curve of human bloodRespiration Physiology, 1980
- Simple, accurate equations for human blood O2 dissociation computationsJournal of Applied Physiology, 1979
- Use of the oxygen electrode in recording Po2 in avian bloodJournal of Applied Physiology, 1968
- Temperature coefficients for Pco2 and Po2 of blood in vitroJournal of Applied Physiology, 1965
- Effect of Temperature on Pco2 and Po2 of Blood in VitroJournal of Applied Physiology, 1956
- THE EFFECT OF TEMPERATURE ON THE pH OF BLOOD AND PLASMA IN VITROJournal of Biological Chemistry, 1948