An elbow extension neuroprosthesis for individuals with tetraplegia
- 1 March 1998
- journal article
- case report
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Rehabilitation Engineering
- Vol. 6 (1) , 1-6
- https://doi.org/10.1109/86.662614
Abstract
Functional electrical stimulation (FES) of the triceps to restore control of elbow extension was integrated into a portable hand grasp neuroprosthesis for use by people with cervical level spinal cord injury. An accelerometer mounted on the upper arm activated triceps stimulation when the arm was raised above a predetermined threshold angle. Elbow posture was controlled by the subjects voluntarily flexing to counteract the stimulated elbow extension. The elbow moments created by the stimulated triceps were at least 4 N/spl middot/m, which was sufficient to extend the arm against gravity. Electrical stimulation of the triceps increased the range of locations and orientations in the workspace over which subjects could grasp and move objects. In addition, object acquisition speed was increased. Thus elbow extension enhances a person's ability to grasp and manipulate objects in an unstructured environment.Keywords
This publication has 14 references indexed in Scilit:
- An externally powered, multichannel, implantable stimulator-telemeter for control of paralyzed muscleIEEE Transactions on Biomedical Engineering, 1998
- Restoration of pronosupination control by FNS in tetraplegia—Experimental and biomechanical evaluation of feasibilityJournal of Biomechanics, 1996
- Orthosis for Improvement of Arm Function in C5/C6 TetraplegiaJPO Journal of Prosthetics and Orthotics, 1996
- A grasp force and position sensor for the quantitative evaluation of neuroprosthetic hand grasp systemsIEEE Transactions on Rehabilitation Engineering, 1995
- A surgically-implanted intramuscular electrode for an implantable neuromuscular stimulation systemIEEE Transactions on Rehabilitation Engineering, 1994
- Instrumented assessment of FNS hand control during specific manipulation tasksIEEE Transactions on Rehabilitation Engineering, 1994
- An analysis of the reliability of percutaneous intramuscular electrodes in upper extremity FNS applicationsIEEE Transactions on Rehabilitation Engineering, 1993
- A multichannel FES system for the restoration of motor functions in high spinal cord injury patients: a respiration-controlled system for multijoint upper extremityIEEE Transactions on Biomedical Engineering, 1989
- Elbow extension in the C5 quadriplegic using functional neuromuscular stimulationIEEE Transactions on Biomedical Engineering, 1989
- Anthropometric Relationships of Body and Body Segment Moments of InertiaPublished by Defense Technical Information Center (DTIC) ,1980