Distinguishing Carotid Artery Pseudo-Occlusion With Color-Flow Doppler

Abstract
Background and Purpose This study was undertaken to determine the impact of color-flow Doppler on the accuracy of noninvasive carotid imaging for distinguishing an internal carotid artery pseudo-occlusion (string sign) from a complete occlusion. Methods From January 1985 to January 1994, review of noninvasive vascular studies, arteriograms, and operative reports of 26 consecutive patients undergoing 27 carotid endarterectomies for carotid pseudo-occlusion was performed. Further review was conducted of all patients identified with carotid occlusion by noninvasive testing who underwent confirmatory arteriography during the same interval. Results Conventional gray-scale duplex scanning (January 1985 to December 1989) correctly identified 3 of 11 (27%) pseudo-occluded internal carotid arteries compared with 15 of 16 (94%) internal carotid artery pseudo-occlusions correctly identified by color-flow Doppler (January 1990 to June 1994) ( P <.01). Similarly, carotid occlusion was more accurately identified by color-flow Doppler (33 of 33, 100%) compared with gray-scale duplex scanning (19 of 27, 90%) ( P <.01). Conclusions The addition of color-flow Doppler to the duplex evaluation of the extracranial carotid circulation improves the accuracy of distinguishing carotid pseudo-occlusion from the occluded internal carotid artery and may obviate the need for arteriography to identify patients with this critical level of carotid stenosis.