Multivariate determination of glucose in whole blood by attenuated total reflection infrared spectroscopy
- 15 September 1989
- journal article
- research article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 61 (18) , 2009-2015
- https://doi.org/10.1021/ac00193a004
Abstract
A spectral analysis of whole EDTA blood was undertaken by using attenuated total reflection and Fourier-transform infrared spectroscopy. The concentration of blood glucose was measured by an enzymatic method using glucose dehydrogenae and ranged between 40 and 290 mg/dL with an average concentration of 90.4 mg/dL. Multivariate calibration with the partial least-squares (PLS) algorithm was performed on spectral data between 1500 and 750 cm-1 showing a varying background from different unidentified interfering compounds. Cross validation was carried out for optimizing the PLS model. PRESS1/2 was 19.8 mg/dL, which was calculated on the basis of 127 standards, whereas the estimated standard deviation for the calibration fit was computed to be 11.9 mg/dL. Infrared spectroscopy can be used for monitoring glucose levels within the normal physiological range in a complex matrix like whole blood as an alternative to electrochemical sensors.Keywords
This publication has 5 references indexed in Scilit:
- Multivariate calibration for assays in clinical chemistry using attenuated total reflection infrared spectra of human blood plasmaAnalytical Chemistry, 1989
- Partial least-squares methods for spectral analyses. 1. Relation to other quantitative calibration methods and the extraction of qualitative informationAnalytical Chemistry, 1988
- Spectroscopic Quantitative Analysis of Strongly Interacting Systems: Human Plasma Protein MixturesApplied Spectroscopy, 1988
- Physicochemical Determinations of Glucose in vivocclm, 1987
- Laser Absorption Spectroscopy with an ATR PrismIEEE Transactions on Biomedical Engineering, 1979