On the field inhomogeneity of a birdcage coil
- 1 September 1994
- journal article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 32 (3) , 418-422
- https://doi.org/10.1002/mrm.1910320319
Abstract
A mathematical model is developed for the numerical analysis of birdcage coils constructed using capacitors having different capacitance. With the aid of this model, it is shown that the deviation of some capacitance introduced by tune, balance, and drive mechanisms can cause inhomogeneity in the B1, field produced by the coil. Useful curves and formulae are developed to provide a guideline for the design of birdcage coils.Keywords
This publication has 2 references indexed in Scilit:
- A technique for double resonant operation of birdcage imaging coilsIEEE Transactions on Medical Imaging, 1989
- Matrix Eigensystem Routines — EISPACK GuideLecture Notes in Computer Science, 1976