Comparison of Rapid Clottable Fibrinogen Assays
Open Access
- 1 November 1972
- journal article
- research article
- Published by Oxford University Press (OUP) in Clinical Chemistry
- Vol. 18 (11) , 1360-1363
- https://doi.org/10.1093/clinchem/18.11.1360
Abstract
Measurements of fibrinogen concentration by use of two rapid methods (clot-weight method and thrombin-turbidity method) were compared with the reference thrombin-coagulation method of Ratnoff and Menzie [J. Lab. Clin. Med. 37, 316 (1951)]. Both rapid methods gave results (in the high, middle, and low ranges) that correlated well with those of the thrombin-coagulation method. However, the thrombin-turbidity method was more precise, easier, and quicker than the clot-weight method.Keywords