One‐Year Follow‐Up of Automatic Adaptation of the Rate Response Algorithm of the QT Sensing, Rate Adaptive Pacemaker
- 1 November 1991
- journal article
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 14 (11) , 1598-1605
- https://doi.org/10.1111/j.1540-8159.1991.tb02735.x
Abstract
Optimal functioning of a rate adaptive pacemaker depends upon reliable sensing of the sensor and appropriate programming of the rate of response algorithm. QT sensing pacemakers use data derived from the endocardial electrogram in the programming of the rate response algorithm. In the latest versions of these pacemakers, programming of the rate response algorithm may be performed using either a semiautomatic Fast Learn (FL) procedure or by using the newly developed, fully Automatic Slope Adaptation (ASA) mechanism. We report our experience in a prospective study of 17 patients in the first year postimplantation. ASA was characterized by significant changes only in the values of the slope settings at the lower rate limit (3.7 msec/msec at time 0 to 5.77 msec/msec at 2 weeks, P less than 0.001) during the first 2 weeks after its enablement. Further adaptation between weeks 2 to 4 was observed (5.77 msec/msec to 6.4 msec/msec, P = 0.2) but this was not significant. The slope settings derived using the FL procedure were also checked at 2 and 4 weeks and were reproducible. They were closest in value to the values attained by the automated mechanism at 4 weeks. This suggests that the final value of the slope setting at the lower rate limit using ASA is reached between weeks 2 to 4. Both methods of slope determination result in satisfactory and similar rate response profiles but the time to achieve slope stability will necessarily be slower with ASA.Keywords
This publication has 9 references indexed in Scilit:
- Usefulness of 1‐Hour and 24‐Hour Heart Rate Holter Inbuilt in New TX* Rate Adaptive PacemakersPacing and Clinical Electrophysiology, 1990
- An office-based exercise protocol for predicting chronotropic response of activity-triggered, rate-variable pacemakersThe American Journal of Cardiology, 1989
- The Ventricular Depolarization Gradient: Effects of Exercise, Pacing Rate, Epinephrine, and Intrinsic Heart Rate Control on the Right Ventricular Evoked ResponsePacing and Clinical Electrophysiology, 1989
- Nonlinear Relationship Between Pacing and Evoked QT IntervalsPacing and Clinical Electrophysiology, 1988
- Reliability of Evoked Endocardial T‐wave Sensing in 1,500 Pacemaker PatientsPacing and Clinical Electrophysiology, 1986
- Reliability of the Evoked Response in Determining the Paced Ventricular Rate and Performance of the QT or Rate Responsive (TX) PacemakerPacing and Clinical Electrophysiology, 1985
- Problems with the Slope Adjustment and Rate Adaptation in Rate Responsive Pacemakers: Oscillation Phenomena and Sudden Rate JumpsPublished by Springer Nature ,1985
- Rate Responsive Pacing — Initial Experience with the QT(TX/Quintech) and Biorate PacemakersPublished by Springer Nature ,1985
- Relation between QT interval and heart rate. New design of physiologically adaptive cardiac pacemaker.Heart, 1981