Rapid protein identification using a microscale electrospray LC/MS system on an ion trap mass spectrometer
- 1 March 1998
- journal article
- Published by American Chemical Society (ACS) in Journal of the American Society for Mass Spectrometry
- Vol. 9 (3) , 194-201
- https://doi.org/10.1016/s1044-0305(97)00282-1
Abstract
A methodology has been developed for the rapid identification of gel separated proteins. Following in gel protein digestion with trypsin, the resulting peptide mixture is analyzed by on-line liquid chromatography, electrospray mass spectrometry (LC/MS). The mass spectral data containing either accurate mass values or sequence specific fragment ion information is then matched to a database of known protein sequences. Key features of the LC/MS system are the use of a novel integrated, microscale LC column-electrospray interface and variable flow solvent delivery to optimize the efficiency of sample loading and gradient elution. With these enhancements, only 10 min is required to analyze each sample. The method is routine for sample amounts ranging from 50 to 500 fmol. The analysis parameters for the ion trap mass spectrometer have to be carefully adjusted in order to keep pace with the rapidly eluting LC peaks. Although designed for rapid LC separations, the integrated column-electrospray interface is also able to provide extended analyses of selected components using a technique known as “peak parking. ”Keywords
This publication has 19 references indexed in Scilit:
- Open-Tubular Capillary Electrochromatography with an On-Line Ion Trap Storage/Reflectron Time-of-Flight Mass Detector for Ultrafast Peptide Mixture AnalysisAnalytical Chemistry, 1997
- Attomole Protein Characterization by Capillary Electrophoresis-Mass SpectrometryScience, 1996
- Femtomole sequencing of proteins from polyacrylamide gels by nano-electrospray mass spectrometryNature, 1996
- Analytical Properties of the Nanoelectrospray Ion SourceAnalytical Chemistry, 1996
- A Microscale Electrospray Interface for Online, Capillary Liquid Chromatography/Tandem Mass Spectrometry of Complex Peptide MixturesAnalytical Chemistry, 1995
- Improvement of an "In-Gel" Digestion Procedure for the Micropreparation of Internal Protein Fragments for Amino Acid SequencingAnalytical Biochemistry, 1995
- Error-Tolerant Identification of Peptides in Sequence Databases by Peptide Sequence TagsAnalytical Chemistry, 1994
- An approach to correlate tandem mass spectral data of peptides with amino acid sequences in a protein databaseJournal of the American Society for Mass Spectrometry, 1994
- Capillary Liquid Chromatography/Mass Spectrometry for Peptide and Protein CharacterizationMethods, 1994