Accurate two‐dimensional IMRT verification using a back‐projection EPID dosimetry method
Top Cited Papers
- 12 January 2006
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 33 (2) , 259-273
- https://doi.org/10.1118/1.2147744
Abstract
The use of electronic portal imaging devices (EPIDs) is a promising method for the dosimetric verification of external beam, megavoltage radiation therapy—both pretreatment and in vivo. In this study, a previously developed EPID back‐projection algorithm was modified for IMRT techniques and applied to an amorphous silicon EPID. By using this back‐projection algorithm, two‐dimensional dose distributions inside a phantom or patient are reconstructed from portal images. The model requires the primary dose component at the position of the EPID. A parametrized description of the lateral scatter within the imager was obtained from measurements with an ionization chamber in a miniphantom. In addition to point dose measurements on the central axis of square fields of different size, we also used dose profiles of those fields as reference input data for our model. This yielded a better description of the lateral scatter within the EPID, which resulted in a higher accuracy in the back‐projected, two‐dimensional dose distributions. The accuracy of our approach was tested for pretreatment verification of a five‐field IMRT plan for the treatment of prostate cancer. Each field had between six and eight segments and was evaluated by comparing the back‐projected, two‐dimensional EPID dose distribution with a film measurement inside a homogeneous slab phantom. For this purpose, the ‐evaluation method was used with a dose‐difference criterion of 2% of dose maximum and a distance‐to‐agreement criterion of . Excellent agreement was found between EPID and film measurements for each field, both in the central part of the beam and in the penumbra and low‐dose regions. It can be concluded that our modified algorithm is able to accurately predict the dose in the midplane of a homogeneous slab phantom. For pretreatment IMRT plan verification, EPID dosimetry is a reliable and potentially fast tool to check the absolute dose in two dimensions inside a phantom for individual IMRT fields. Film measurements inside a phantom can therefore be replaced by EPID measurements.Keywords
Funding Information
- KWF Kankerbestrijding (NKI 2000-2255)
This publication has 47 references indexed in Scilit:
- The long‐term stability of amorphous silicon flat panel imaging devices for dosimetry purposesMedical Physics, 2004
- Verification of treatment parameter transfer by means of electronic portal dosimetryMedical Physics, 2004
- Dosimetric properties of an amorphous silicon electronic portal imaging device for verification of dynamic intensity modulated radiation therapyMedical Physics, 2003
- Dosimetric investigation and portal dose image prediction using an amorphous silicon electronic portal imaging deviceMedical Physics, 2001
- Direct measurement and analytical modeling of scatter in portal imagingMedical Physics, 2000
- Dosimetric characteristics of a liquid-filled electronic portal imaging deviceInternational Journal of Radiation Oncology*Biology*Physics, 1995
- Portal dose measurement in radiotherapy using an electronic portal imaging device (EPID)Physics in Medicine & Biology, 1995
- Daily dosimetric quality control of the MM50 Racetrack Microtron using an electronic portal imaging deviceRadiotherapy and Oncology, 1995
- The design and performance characteristics of a multileaf collimatorPhysics in Medicine & Biology, 1994
- The determination of phantom and collimator scatter components of the output of megavoltage photon beams: measurement of the collimator scatter part with a beam-coaxial narrow cylindrical phantomRadiotherapy and Oncology, 1991