Evaluation of the HiFocus® electrode array with positioner in human temporal bones
- 1 July 2003
- journal article
- research article
- Published by Cambridge University Press (CUP) in The Journal of Laryngology & Otology
- Vol. 117 (7) , 527-531
- https://doi.org/10.1258/002221503322112932
Abstract
The aim of the study presented was to assess the insertion mode and possible intracochlear trauma after implantation of the HiFocus® electrode with positioner in human temporal bones. The study was performed in five freshly frozen temporal bones. The position of electrodes was evaluated using conventional X-ray analysis, rotational tomography and histomorphological analysis. Insertion of the HiFocus® electrode with positioner resulted in considerable trauma to fine cochlear structures including fracture of the osseous spiral lamina, dislocation of the electrode array from the scala tympani into the scala vestibuli and fracture of the modiolus close to the cochleostomy. The implication of the results regarding clinical outcome will be discussed.Keywords