Accuracy of Devices Used for Self-Monitoring of Blood Glucose
Open Access
- 1 January 1998
- journal article
- research article
- Published by SAGE Publications in Annals of Clinical Biochemistry: International Journal of Laboratory Medicine
- Vol. 35 (1) , 68-74
- https://doi.org/10.1177/000456329803500108
Abstract
We measured the inaccuracy of 17 home blood glucose monitors (two visually read, eight colorimetric and seven amperometric). Using strips from a single batch, blood glucose measurements were performed by three medical laboratory technologists on at least 50 capillary blood specimens from patients attending two diabetes clinics. Additional capillary blood was deproteinized and assayed with hexokinase to give a whole blood glucose result. A dedicated glucose analyser was also studied to cross-validate the methodology. At a mean glucose concentration of about 9 mmol/L, monitor readings differed from the reference results by −5.1 to + 19.5% with three systems failing to meet the American Diabetes Association guideline for total error of less than 15%. This problem would be alleviated by manufacturers adopting a common policy on calibration and on reporting as a plasma or whole blood value.Keywords
This publication has 9 references indexed in Scilit:
- The Yellow Springs Analyser Gives Lower Results than other Glucose Oxidase Methods of Glucose MeasurementAnnals of Clinical Biochemistry: International Journal of Laboratory Medicine, 1995
- Monitoring Blood Glucose MetersAmerican Journal of Clinical Pathology, 1995
- Whole Blood Glucose: What are We Actually Measuring?Annals of Clinical Biochemistry: International Journal of Laboratory Medicine, 1995
- Self-Monitoring of Blood GlucoseDiabetes Care, 1994
- The Effect of Intensive Treatment of Diabetes on the Development and Progression of Long-Term Complications in Insulin-Dependent Diabetes MellitusNew England Journal of Medicine, 1993
- What is Blood Glucose: Can it be Measured?Diabetic Medicine, 1990
- Consensus Statement on Self-Monitor ing of Blood GlucoseDiabetes Care, 1987
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- Evaluation of a New Blood Glucose Monitoring System: Importance of the Appropriate Reference MethodDiabetes Care, 1984