Confocal scanning light microscopy with high aperture immersion lenses
- 1 November 1979
- journal article
- research article
- Published by Wiley in Journal of Microscopy
- Vol. 117 (2) , 219-232
- https://doi.org/10.1111/j.1365-2818.1979.tb01178.x
Abstract
SUMMARY: The imaging characteristics of a confocal scanning light microscope (CSLM) with high aperture, immersion type, lenses (N.A. = 1·3) are investigated. In the confocal arrangement the images of the illumination and detector pinholes are made to coincide in a common point, through which the object is scanned mechanically. Results show that for point objects the theoretically expected improved response by a factor of 1·4 in comparison with standard microscopy can indeed be realized. Low side lobe intensity and absence of glare permits the imaging at high resolution of weak details close to strong features. A further improvement by a factor of 1·25 in point resolution in CSLM is found after apodization with an annular aperture. Due to the scanning approach all possibilities of electronic image processing become available in light microscopy.This publication has 6 references indexed in Scilit:
- Imaging modes in confocal scanning light microscopy (CSLM)Journal of Microscopy, 1979
- Confocal scanning light microscopy with high aperture immersion lensesJournal of Microscopy, 1979
- Image Formation in the Scanning MicroscopeOptica Acta: International Journal of Optics, 1977
- Acoustic microscope—scanning versionApplied Physics Letters, 1974
- The Extrapolating Pupil, Image Synthesis, and Some Thought ApplicationsApplied Optics, 1970
- The airy disc formula for systems of high relative apertureProceedings of the Physical Society, 1943