Complex bioelectric impedance measurement system for the frequency range from 5 Hz to 1 MHz
- 1 March 1993
- journal article
- Published by Springer Nature in Annals of Biomedical Engineering
- Vol. 21 (2) , 135-146
- https://doi.org/10.1007/bf02367609
Abstract
Analytic techniques that have been successfully employed in materials science, and to a lesser extent in the study of biologic systems, have potential for improving the application of bioelectric impedance provided that both real and imaginary impedance components can be measured with sufficient accuracy over a given frequency range. Since biologic tissue, particularly animal tissue, is typically highly conductive, phase angles are small, making accurate measurements difficult. A practical four-terminal system employing commercial lock-in amplifiers is described and error sources and corrective tech-niques are discussed.Keywords
This publication has 2 references indexed in Scilit:
- Methods of complex impedance measurements in biologic tissue.1984
- Reduction of Interference Due to Common Mode Voltage in Biopotential AmplifiersIEEE Transactions on Biomedical Engineering, 1983