Instrumental Infrared Estimation of Fat, Protein, and Lactose in Milk: Collaborative Study
- 1 September 1978
- journal article
- research article
- Published by Oxford University Press (OUP) in Journal of AOAC INTERNATIONAL
- Vol. 61 (5) , 1015-1034
- https://doi.org/10.1093/jaoac/61.5.1015
Abstract
Thirty-six milks pre-analyzed by accepted standard methods were analyzed with Milkoscan instruments at 6 laboratories. Using pairs comparisons as described by Youden, precision errors for fat were about 0.033% for the first and 0.022% for the second of duplicate tests, indicating a carry-over effect for the first tests. Precision errors for protein were about 0.021 % for both tests. Systematic errors for fat were about 0.17% with calibrations based on reference analyses at each laboratory but were reduced to about 0.044% with calibrations by reference analyses from one laboratory. With homogenized milks, systematic errors for fat increased to about 0.07% and were larger at high fat levels. This effect could be minimized by changing the wavelength of maximum transmission for the fat filter. Systematic errors of about 0.067% for protein were reduced to about 0.03% with calibrations based on common reference analyses. Satisfactory calibrations for lactose were obtained with the 2 Milkoscan 203 models with standard errors of estimate of 0.034 and 0.033%. Results for homogenized and unhomogenized milk were diferent for most instruments, indicating the need to calibrate for them separately.This publication has 0 references indexed in Scilit: