Temperature Monitoring During Radiofrequency Ablation
- 1 February 1996
- journal article
- review article
- Published by Wiley in Journal of Cardiovascular Electrophysiology
- Vol. 7 (2) , 163-173
- https://doi.org/10.1111/j.1540-8167.1996.tb00511.x
Abstract
Temperature Monitoring During RF Ablation. Thermal injury is the primary mechanism of lesion formation during radiofrequency catheter ablation procedures. Irreversible tissue injury requires heating to approximately 50°C. Temperatures above 100°C result in coagulum formation. Because of this importance of temperature during radiofrequency catheter ablation procedures, temperature monitoring has been proposed as a tool to facilitate catheter ablation procedures. The results of recent clinical studies demonstrate that electrode temperatures do not differ at successful and failed ablation sites, electrode temperature does not predict or eliminate the possibility of arrhythmia recurrence, and closed‐loop temperature control decreases but does not eliminate the development of coagulum nor guarantees that target temperatures will be achieved. These observations are due in large part to the important distinctions between electrode temperature, the temperature at the electrode‐tissue interface, and the temperature at the ablation target. Nonetheless, temperature monitoring and temperature control are valuable tools during radiofrequency ablation procedures as they provide important information regarding the adequacy of tissue heating, minimize the development of coagulum, and maximize lesion size.Keywords
This publication has 19 references indexed in Scilit:
- 1008-11 Recurrence of Conduction Following Catheter Ablation Procedures: Relationship to Electrode TemperaturesJournal of the American College of Cardiology, 1995
- Effect of accessory pathway location on the efficiency of heating during radiofrequency catheter ablationAmerican Heart Journal, 1995
- Cellular electrophysiological effects of hyperthermia on isolated guinea pig papillary muscle. Implications for catheter ablation.Circulation, 1993
- Temperature-controlled radiofrequency catheter ablation of AV conduction: first clinical experienceEuropean Heart Journal, 1993
- Temperature monitoring during radiofrequency catheter ablation of accessory pathways.Circulation, 1992
- Assessment of Effects of a Radiofrequency Energy Field and Thermistor Location in an Electrode Catheter on the Accuracy of Temperature MeasurementPacing and Clinical Electrophysiology, 1991
- Observations on electrode-tissue interface temperature and effect on electrical impedance during radiofrequency ablation of ventricular myocardium.Circulation, 1990
- Catheter ablation of the atrioventricular junction with radiofrequency energy.Circulation, 1989
- Tissue Heating During Radiofrequency Catheter Ablation: A Thermodynamic Model and Observations in Isolated Perfused and Superfused Canine Right Ventricular Free WallPacing and Clinical Electrophysiology, 1989
- Controlled Destruction and Temperature Distributions in Biological Tissues Subjected to Monoactive ElectrocoagulationJournal of Biomechanical Engineering, 1980