Can a Commercial Diagnostic Ultrasound Device Accelerate Thrombolysis?

Abstract
Background and Purpose— Recently, 3 clinical trials revealed encouraging results in recanalization and clinical outcome in acute stroke patients when 2-MHz transcranial Doppler monitoring was applied. This study investigated whether a 1.8-MHz commercial diagnostic ultrasound device has the potential to facilitate thrombolysis using an in vitro stroke model. Methods— Duplex-Doppler, continuous wave-Doppler, and pulsed wave (PW)-Doppler were compared on their impact on recombinant tissue plasminogen activator (rtPA)–mediated thrombolysis. Blood clots were transtemporally sonicated in a human stroke model. Furthermore, ultrasound attenuation of 5 temporal bones of different thickness was determined. Results— In comparison, only PW-Doppler accelerated rtPA–mediated thrombolysis significantly. Without temporal bone, PW-Doppler plus rtPA showed a significant enhancement in relative clot weight loss of 23.7% when compared with clots treated with rtPA only (33.9±5.5% versus 27.4±5.2%; P Conclusion— Without temporal bone, PW-Doppler significantly enhanced thrombolysis. However, because of a high attenuation of ultrasound by temporal bone, no thrombolytic effect was observed in our in vitro model, although Doppler imaging through the same temporal bone was still possible.