Equipment deadspace and drug administration
Open Access
- 22 February 1990
- journal article
- Published by Wiley in Anaesthesia
- Vol. 45 (2) , 145-147
- https://doi.org/10.1111/j.1365-2044.1990.tb14283.x
Abstract
Equipment deadspace can be a hazard in drug administration. The aim of this study was to measure the size of this problem. The deadspaces of a selection of cannulae, syringes and epidural catheters were measured by displacing the deadspace volume with water and measuring the weight gain. Between 10 and 30% of a 1-ml intravenous dose will remain in the deadspace depending on the type and gauge of the cannulae. The deadspace of an epidural catheter and filter is approximately 1 ml. All cannulae should be flushed after drug administration. Epidural catheters must be flushed with the same solution as the test dose before insertion.Keywords