Some consideration of data corrections for a SAXS system employing a linear PSPC
- 1 August 1989
- journal article
- Published by International Union of Crystallography (IUCr) in Journal of Applied Crystallography
- Vol. 22 (4) , 376-378
- https://doi.org/10.1107/s0021889889003195
Abstract
When a one-dimensional position-sensitive proportional counter (PSPC) is employed in a small-angle X-ray scattering (SAXS) system, detector-related problems might be encountered, such as nonuniformity of the detector, resolution function defined by the geometry of the receiving slit etc., all of which could cause distortion to the measured intensity profile. A set of experiments has been performed to test the significance of some of these problems, and corrections have been worked out to circumvent them. To obtain an optimized condition between flux and resolution, a wedge-shaped receiving slit is recommended.Keywords