Optimal pitch in spiral computed tomography
- 1 October 1997
- journal article
- Published by Wiley in Medical Physics
- Vol. 24 (10) , 1635-1639
- https://doi.org/10.1118/1.597971
Abstract
The focus of this paper is to analytically optimize spiral/helical computed tomography (CT) protocols based on a simplified imaging model. Spiral CT was approximately modeled as follows: Using the half‐scan raw data interpolation method, the variance of the spiral CT slice sensitivity profile is equal to the sum of squared detector collimation divided by 12 and squared table increment divided by 24. Image noise variance is inversely proportional to tube current and detector collimation. The maximum continuous scanning time is inversely proportional to tube current. Slice thickness, image noise, and signal‐to‐noise ratio were, respectively, optimized for a given scanning coverage, consistently resulting in pitch of square root of 2. To avoid longitudinal aliasing, at least 2–3 transverse slices should be reconstructed per collimation. When the simplified spiral CT model is valid and a scanning range specified, 1.4 pitch is required for optimal image quality. The method can be applied to more accurate spiral CT models.Keywords
This publication has 12 references indexed in Scilit:
- Low-contrast resolution in volumetric x-ray CT-Analytical comparison between conventional and spiral CTMedical Physics, 1997
- Maximum volume coverage in spiral computed tomography scanningAcademic Radiology, 1996
- Increased Scan Pitch for Vascular and Thoracic Spiral CTRadiology, 1995
- Spatial variation of section sensitivity profile in spiral computed tomographyMedical Physics, 1994
- Stair-step artifacts in three-dimensional helical CT: an experimental study.Radiology, 1994
- Longitudinal resolution in volumetric x-ray computerized tomography-Analytical comparison between conventional and helical computerized tomographyMedical Physics, 1994
- Spiral CT of the lungs: optimal technique and resolution compared with conventional CT.American Journal of Roentgenology, 1994
- Helical CT image noise-analytical resultsMedical Physics, 1993
- Evaluation of section sensitivity profiles and image noise in spiral CT.Radiology, 1992
- Computed tomography scanning with simultaneous patient translationMedical Physics, 1990