Improved method for measurement of inorganic phosphate in serum with a centrifugal analyzer.

Abstract
A direct mehtod [Clin. Chim. Acta 46, 113 (1973)] for determination of inorganic phosphate in serum was adapted for use with a centrifugal analyzer. Contamination is minimized and analysis rate maximized by doing the reaction in the reagent wells of the transfer disc and by utilizing the high-speed spectrophotometric and data-reduction capabilities of the centrifugal analyzer. Hemolysis, icterus, and moderate lipemia cause no interference. Grossly lipemic sera and sera from patients with plasma cell dyscrasias can be analyzed by incorporating appropriate blanking and dilution techniques. The method exhibits excellent sensitivity and precision and results correlate well with those from a continuous-flow procedure.

This publication has 0 references indexed in Scilit: