Validation of saliva sampling for total body water determination by H 2 18 O dilution
Open Access
- 1 March 1982
- journal article
- research article
- Published by Elsevier in The American Journal of Clinical Nutrition
- Vol. 35 (3) , 591-594
- https://doi.org/10.1093/ajcn/35.3.591
Abstract
The use of saliva sampling for the measurement of total body water by H218O stable isotope dilution was assessed. The H218O abundance in saliva reached a constant value by three hours after isotope administration, and the total body water calculated from the 3-h saliva values was not significantly different from the value calculated from the 4- or 6-h serum samples. The results indicated that isotopic equilibration between these fluids was rapid and complete. Comparison of total body water values calculated from the paired serum and saliva studies indicated that the relative precision of the determination was between 0.8 and 1.5%. Saliva sampling is noninvasive and is well suited for use in children and in field studies.This publication has 6 references indexed in Scilit:
- Total body water measurement in humans with 18O and 2H labeled waterThe American Journal of Clinical Nutrition, 1980
- Oxygen and Hydrogen Isotopic Ratios in Plant CelluloseScience, 1977
- Precise measurement of total body water using trace quantities of deuterium oxideJournal of Mass Spectrometry, 1977
- The measuring procedure and corrections for the high-precision mass-spectrometric analysis of isotopic abundance ratios, especially referring to carbon, oxygen and nitrogenInternational Journal of Mass Spectrometry and Ion Physics, 1973
- Isotopic standards for carbon and oxygen and correction factors for mass-spectrometric analysis of carbon dioxideGeochimica et Cosmochimica Acta, 1957
- Measurement of Total Carbon Dioxide Production by Means of D2O18Journal of Applied Physiology, 1955