Enlargement of basilar artery aneurysms following balloon occlusion — “water-hammer effect”
- 1 December 1991
- journal article
- case report
- Published by Journal of Neurosurgery Publishing Group (JNSPG) in Journal of Neurosurgery
- Vol. 75 (6) , 963-968
- https://doi.org/10.3171/jns.1991.75.6.0963
Abstract
✓ Two patients with distal basilar aneurysms were treated with intra-aneurysmal balloon occlusion. After apparently successful therapy, follow-up angiograms demonstrated aneurysm enlargement with balloon migration distally in the sac. Geometric mismatch between the base of the balloons and the aneurysm neck together with transmitted pulsation through the 2-hydroxyl-ethylmethacrylate (HEMA)-filled balloon directly contributed to aneurysm enlargement. In this report, the authors discuss the problems of progressive aneurysm enlargement due to a “water-hammer effect” and the possibility of hemorrhage following subtotal occlusion.Keywords
This publication has 8 references indexed in Scilit:
- Rupture of aneurysms following balloon embolizationJournal of Neurosurgery, 1990
- Endovascular treatment of intracranial aneurysms with a new silicone microballoon device: technical considerations and indications for therapy.Radiology, 1990
- Detachable balloon embolization therapy of posterior circulation intracranial aneurysmsJournal of Neurosurgery, 1989
- Regrowth of aneurysm sacs from residual neck following aneurysm clippingJournal of Neurosurgery, 1989
- Permanent inflation of detachable balloons with a low-viscosity, hydrophilic polymerizing system.Radiology, 1988
- Balloon embolization of a large distal basilar artery aneurysmJournal of Neurosurgery, 1986
- Clinicopathological study of cerebral aneurysmsJournal of Neurosurgery, 1978
- Physical factors in the initiation, growth, and rupture of human intracranial saccular aneurysmsJournal of Neurosurgery, 1972