A Method for Noninvasive Determination of Inspiratory Resistance during Proportional Assist Ventilation
- 15 March 2001
- journal article
- research article
- Published by American Thoracic Society in American Journal of Respiratory and Critical Care Medicine
- Vol. 163 (4) , 829-839
- https://doi.org/10.1164/ajrccm.163.4.2005063
Abstract
Currently available noninvasive methods for measuring inspiratory resistance (RI) are difficult to implement or interpret during assisted mechanical ventilation on account of the confounding effect of respiratory efforts (Pmus). We propose a simple method consisting of brief reductions in airway pressure (Paw) in the early part of the inflation phase (pulse). Paw, flow (V), and volume (V) are measured at the beginning of the pulse (T (0)), at the trough of the pulse (TI) and at a point 0.1 s before T(0) (T(-1)). Equations of motion of the form [Pmus + Paw = V. K(1) + V (2). K(2) +V. E] are generated for the data at the three time points (E = elastance, K(1) and K(2) are Rohrer's constants). These three equations can be solved for K(1) and K(2) if it is arranged that the pulse has appropriate configuration and timing, and if it is assumed that DeltaPmus/Deltat is constant over the brief pulse period. The method was tested in 67 patients ventilated with proportional assist ventilation (PAV). The results were compared with those obtained using the interrupter technique during a period of controlled mechanical ventilation (CMV). RI, expressed at a standard flow of 1 L. s(-)(1), was slightly higher during PAV (16.4 +/- 4.9 versus 15.5 +/- 4.5 cm H(2)O. L(-1). s, p < 0.001). The average difference was 0.9 +/- 2.0 cm H(2)O. L(-1). s, corresponding to 5.4 +/- 12.6% of the average of RCMV and RPAV. The correlation coefficient was 0.92 (p = 8E-28) with a slope (1.01) and intercept (0.8) not significantly different from 1.0 and 0, respectively. We conclude that brief negative pulses applied early during the inflation phase can be used to provide reliable estimates of inspiratory resistance during PAV.Keywords
This publication has 34 references indexed in Scilit:
- A simplified method for monitoring respiratory impedance during continuous positive airway pressureEuropean Respiratory Journal, 2000
- Detection of flow limitation during tidal breathing by the interruptor techniqueEuropean Respiratory Journal, 1995
- Servocontrolled generator to measure respiratory impedance from 0.25 to 26 Hz in ventilated patients at different PEEP levelsEuropean Respiratory Journal, 1995
- Proportional Assist Ventilation: Results of an Initial Clinical TrialAmerican Review of Respiratory Disease, 1992
- Proportional Assist Ventilation, a New Approach to Ventilatory Support: TheoryAmerican Review of Respiratory Disease, 1992
- Evidence for reflex upper airway dilator muscle activation by sudden negative airway pressure in man.The Journal of Physiology, 1991
- The Ventilatory Recruitment Threshold for Carbon DioxideAmerican Review of Respiratory Disease, 1990
- In VitroversusIn VivoComparison of Endotracheal Tube Airflow ResistanceAmerican Review of Respiratory Disease, 1989
- Measurement of respiratory mechanics using the siemens servo ventilator 900CPediatric Pulmonology, 1987
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986