Detection Efficiency of a Continuous Dynode Electron Multiplier (Ceratron) for Positrons
- 1 May 1982
- journal article
- Published by IOP Publishing in Japanese Journal of Applied Physics
- Vol. 21 (5R) , 702-705
- https://doi.org/10.1143/jjap.21.702
Abstract
The detection efficiency of the continuous dynode electron multiplier, Ceratron, for positrons has been obtained as a function of the incident energy E in for a slow positron beam measurement. For comparison, the efficiency curve for a channel electron multiplier (CEM) was also obtained under the same conditions. In the use of a Ceratron, the counting rate in conventional use is low compared with that under the best conditions, obtained using a new method. The dependence of the efficiency on E in is very large for the Ceratron, but rather small for the CEM. The dependences in the low energy region for the Ceratron and for the CEM show that calibration of the efficiency on E in for each detector is necessary in the conventional use for electrons. The detection efficiency of a Ceratron for positrons under the best conditions is almost the same as that for a CEM.Keywords
This publication has 10 references indexed in Scilit:
- Low energy positrons from metal surfacesSurface and Interface Analysis, 1980
- Ceramic Secondary Electron Multipliers for Detecting Charged Paticles and RadiationTransactions of the Society of Instrument and Control Engineers, 1975
- Ceramic Secondary Electron MultiplierSHINKU, 1974
- The Calibration of Electrostatic Analyzers and Channel Electron Multipliers Using Laboratory Simulated Omnidirectional Electron BeamsReview of Scientific Instruments, 1973
- Channel electron multiplier efficiency for 10–1000 eV electronsNuclear Instruments and Methods, 1971
- Absolute Efficiency Measurements for Channel Electron Multipliers Utilizing a Unique Electron SourceReview of Scientific Instruments, 1970
- Degradation of Continuous-Channel Electron Multipliers in a Laboratory Operating EnvironmentReview of Scientific Instruments, 1969
- Channeltron Fatigue and Efficiency for Protons and ElectronsReview of Scientific Instruments, 1969
- A Directional, Low Energy Electron Detector Employing Channel Electron MultipliersIEEE Transactions on Nuclear Science, 1968
- Caracteristiques des multiplicateurs d'electrons sans fenetre; Utilisation en fusee comme detecteurs d'electrons auroraux de faible energieNuclear Instruments and Methods, 1967