Automated On-line Preconcentration for Trace Metals Determination in Water Samples by Inductively Coupled Plasma Mass Spectrometry
- 1 January 1997
- journal article
- research article
- Published by Royal Society of Chemistry (RSC) in Journal of Analytical Atomic Spectrometry
- Vol. 12 (5) , 563-566
- https://doi.org/10.1039/a607251h
Abstract
A preconcentration flow system using the cationic resin AG50W-X8 for multielement determinations of trace constituents in water samples to work on-line with ICP-MS is described. The flow network comprises two small resin columns inserted in parallel channels of the injector port. Sample loading onto the columns occurs simultaneously and independently. Elution takes place intermittently through each column, releasing eluent fractions containing similar concentrations of the analytes, which are merged downstream. The volume of eluent in the alternate aliquots sent through the columns was adjusted to attain precise transient signals. Enrichment factors from 10 to 95 for sample loading times of 0.5 to 5 min were attained. The eluted band produced a transient peak with width at half peak maximum of approximately 20 s, enough to scan several masses for multielement determinations by ICP-MS. The feasibility of the proposed system was demonstrated by determination of Cd, Pb, Ni, Cu and Zn in the certified reference water sample SLRS-1. Accurate results were obtained using a preconcentration factor of 30 fold and a sample throughput of 20 samples per hour.Keywords
This publication has 12 references indexed in Scilit:
- Multicommutation in flow analysis. Part 1. Binary sampling: concepts, instrumentation and spectrophotometric determination of iron in plant digestsAnalytica Chimica Acta, 1994
- Determination of trace metals in sea-water and the on-line removal of matrix interferences by flow injection with inductively coupled plasma mass spectrometric detectionJournal of Analytical Atomic Spectrometry, 1994
- On-line preconcentration system for inductively coupled plasma atomic emission spectrometry with quinolin-8-ol and Amberlite XAD-2 resinAnalytica Chimica Acta, 1992
- Determination of cadmium, lead and copper in water samples by flame atomic-absorption spectrometry with preconcentration by flow-injection on-line sorbent extractionTalanta, 1991
- Determination of ultra-trace levels of metal ions in sea-water with on-line pre-concentration and electrothermal atomic absorption spectrometryJournal of Analytical Atomic Spectrometry, 1991
- Determination of trace metals in reference water standards by inductively coupled plasma mass spectrometry with on-line preconcentrationAnalytical Chemistry, 1989
- Performance studies under flow conditions of silica-immobilized 8-quinolinol and its application as a preconcentration tool in flow injection/atomic absorption determinationsAnalytical Chemistry, 1985
- An efficient flow-injection system with on-line ion-exchange preconcentration for the determination of trace amounts of heavy metals by atomic absorption spectrometryAnalytica Chimica Acta, 1984
- Combination of flow injection analysis with flame atomic-absorption spectrophotometry: determination of trace amounts of heavy metals in polluted seawaterThe Analyst, 1983
- Ion exchange in flow injection analysis: Determination of Ammonium Ions at the μg 1-1 level in Natural Waters with Pulsed Nessler ReagentAnalytica Chimica Acta, 1980