CONSTRUCTION AND USE OF A SEMIMICRO LIGHT-SCATTERING APPARATUS
- 1 November 1953
- journal article
- Published by Canadian Science Publishing in Canadian Journal of Chemistry
- Vol. 31 (11) , 1078-1092
- https://doi.org/10.1139/v53-141
Abstract
A light-scattering apparatus requiring only 4 ml. of liquid was designed and built. The device has permitted the measurement of intensity for angles between [Formula: see text] and 135° to the incident beam. In this range the unpolarized, horizontal, and vertical components of the light emitted from a solution of fluorescein were constant to within ± 0.2%, 0.4%, and 0.6% respectively. The apparatus was calibrated with a suspension of Ludox colloidal silica which absorbed no light for λ = 3800 Å to 6000 Å. The suspension was found to possess all the properties of a Rayleigh scattering system except randomness. This condition was achieved by extrapolating the light-scattering and spectrophotometric functions to zero concentration when computing the calibration factor. The molecular weight of a sample of sodium carrageenate, determined by the method of Zimm, was 1,700,000. The distribution of intensity at. zero concentration fell between that for stiff rods and polydisperse coils.Keywords
This publication has 0 references indexed in Scilit: