Speckle Reduction in Pulse Echo Imaging Using Phase Insensitive and Phase Sensitive Signal Processing Techniques
- 1 January 1986
- journal article
- research article
- Published by SAGE Publications in Ultrasonic Imaging
- Vol. 8 (1) , 11-28
- https://doi.org/10.1177/016173468600800102
Abstract
In conventional B-scan imaging, speckle often distorts the true representation of small structures and lesions and hence reduces the diagnostic potential of the technique. Considerable speckle reduction has been achieved with an f/2.4, cone hybrid system by cutting the large aperture transducers into sectors, and using either phase insensitive sector addition (PISA), phase insensitive sector multiplication (PISM), or an rf multiplicative processing technique. We have compared two modes of operation with the hybrid system. In the first, the cone acts as transmitter, and each sector of the f/2.4 spherical aperture is a receiver. In the second mode, the f/2.4 sphere acts as transmitter, and each sector of the cone is a receiver. Quantitative assessment using the contrast to speckle ratio developed by Patterson et al, indicates that processing under both modes results in reduced speckle noise and improved image quality, with the cone sector receive (rx) geometry offering the greatest improvement.Keywords
This publication has 18 references indexed in Scilit:
- Quantitative contrast measurements in B-mode images comparison between experiment and theoryUltrasound in Medicine & Biology, 1986
- Computer Simulations of Speckle in B-Scan ImagesUltrasonic Imaging, 1983
- The Improvement and Quantitative Assessment of B-Mode Images Produced by an Annular Array/Cone HybridUltrasonic Imaging, 1983
- An Annular Array System for High Resolution Breast EchographyUltrasonic Imaging, 1982
- The Conical Scanner: A Two Transducer Ultrasound Scatter Imaging TechniqueUltrasonic Imaging, 1981
- Acoustic Speckle: Theory and Experimental AnalysisUltrasonic Imaging, 1979
- Annular array design and logarithmic processing for ultrasonic imagingUltrasound in Medicine & Biology, 1978
- High-Resolution B-Scan Systems Using a Circular ArrayPublished by Springer Nature ,1975
- Acoustic Imaging with Thin Annular AperturesPublished by Springer Nature ,1974
- Ultrasound Axicon: a device for focusing over a large depthThe Journal of the Acoustical Society of America, 1973