Detection of pulmonary nodules by multislice computed tomography: improved detection rate with reduced slice thickness
Top Cited Papers
- 1 October 2003
- journal article
- research article
- Published by Springer Nature in European Radiology
- Vol. 13 (10) , 2378-2383
- https://doi.org/10.1007/s00330-003-1915-7
Abstract
The purpose of this study was to find out if the use of 1.25-mm collimated thin-slice technique helps to detect more small pulmonary lung nodules than the use of 5 mm. A total of 100 patient examinations that allowed a reconstruction of 1.25-mm slice thickness in addition to the standard of 5-mm slices were included in a prospective study. Acquisition technique included four rows of 1-mm slices. Two sets of contiguous images were reconstructed and compared with 1.25- and 5-mm slice thickness, respectively. Two radiologists performed a film-based analysis of the images. The size and the confidence of the seen nodules were reported. We did not perform a histological verification, according to the normal clinical procedure, although it would be optimal regarding research. Statistical analysis was performed by using longitudinal analysis described by Brunner and Langer [10]. In addition, sensitivity, specificity, negative predictive value and positive predictive value were calculated for each reader using the 1.25-mm sections as the gold standard. As an index for concordance the kappa value was used. A value of pp<0.05). Using 1.25 mm as the gold standard, sensitivity for 5-mm reconstruction interval was 88 and 86% for observers A and B, respectively. No false-positive results were reported for 5-mm sections. Interobserver agreement for nodule detection determined for 1.25-mm reconstruction intervals showed a k value of 0.753, indicating a good agreement, and 0.562 for 5-mm reconstruction intervals, indicating a moderate agreement. Brunner and Langer [10] analysis showed significant differences for slice thickness and no significant difference between the observers. Reduced slice thickness demonstrated an improvement of small nodule detection, confidence levels, and interobserver agreement. Application of thin-slice multidetector-row CT may raise the sensitivity for lung nodule detection, although the higher detection rate of smaller nodules has to be evaluated from a clinical perspective and remains problematic about how the detection of small nodules will effect patient outcome.Keywords
This publication has 21 references indexed in Scilit:
- Screening for Early Lung Cancer with Low-Dose Spiral CT: Prevalence in 817 Asymptomatic SmokersRadiology, 2002
- Lung cancer screening by low-dose spiral computed tomographyEuropean Respiratory Journal, 2001
- Mehrschicht-Spiral-CT der Lunge: Multiplanare Rekonstruktionen und Maximum-Intensitäts-Projektionen in der Detektion von LungenrundherdenRöFo - Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebenden Verfahren, 2001
- Early Lung Cancer Action Project: overall design and findings from baseline screeningThe Lancet, 1999
- Cancer statistics, 1999CA: A Cancer Journal for Clinicians, 1999
- Glossary of terms for CT of the lungs: recommendations of the Nomenclature Committee of the Fleischner Society.Radiology, 1996
- Determining the likelihood of malignancy in solitary pulmonary nodules with Bayesian analysis. Part I. Theory.Radiology, 1993
- Pulmonary nodule: evaluation with spiral volumetric CT.Radiology, 1991
- Comparison of whole lung tomography and computed tomography for detecting pulmonary nodulesAmerican Journal of Roentgenology, 1978
- 1976 Caldwell Lecture: varying manifestation of peripheral pulmonary neoplasms: a radiologic-pathologic correlative studyAmerican Journal of Roentgenology, 1977