Continuous measurement of minute ventilation and gaseous metabolism of newborn infants
- 1 May 1981
- journal article
- research article
- Published by American Physiological Society in Journal of Applied Physiology
- Vol. 50 (5) , 1098-1103
- https://doi.org/10.1152/jappl.1981.50.5.1098
Abstract
A system of instrumentation for continuous measurement of gaseous metabolism and minute volume (VI) in the human newborn is described. O2 uptake and CO2 production are measured by open-circuit techniques utilizing a Servomex OA184 differential paramagnetic O2 analyzer and a BEckman LB-2 infrared CO2 analyzer. VI is measured with bias-flow pneumotachometry. Bench performance is described, methodological errors are defined, and clinical data are presented. The instrumentation is capable of safe, accurate, and continuous measurement of respiratory and metabolic variables in low-birth-weight infants.This publication has 0 references indexed in Scilit: