Fetal monitoring with pulse oximetry
- 1 January 1991
- journal article
- Published by Wiley in BJOG: An International Journal of Obstetrics and Gynaecology
- Vol. 98 (1) , 36-41
- https://doi.org/10.1111/j.1471-0528.1991.tb10308.x
Abstract
Summary. Continuous fetal monitoring was achieved with a fetal scalp pulse oximetry sensor in 86 labours. The average recorded fetal oxygen saturation in early labour (cervical dilatation < 5 cm) was 68% (SD 13%). At the end of labour (cervical dilatation ≥ 9 cm) the recorded mean oxygen saturation was 58% (SD 17%). The largest range of readings during a single labour was 81%-11% but this drop was associated with cord compression. The average SD during 1 h of normal labour was 10%. A second group of 40 fetuses was monitored during induction of labour before and after elective amniotomy. Oxygen saturation did not appear to change after amniotomy (mean change −0.4%, SD 1.2%) and there was no difference between mean antenatal or early intrapartum readings. We excluded the amniochorionic membranes as a possible source of data corruption by measuring their in vitro absorption spectra and confirming that they do not preferentially absorb light of either 660 or 940 nm wavelength. Non-invasive pulse oximetry can be used to monitor the fetus before and during labour.Keywords
This publication has 16 references indexed in Scilit:
- The effect of caput succedaneum on oxygen saturation measurementsBJOG: An International Journal of Obstetrics and Gynaecology, 1990
- The effect of meconium on neonatal and fetal reflectance pulse oximetryjpme, 1990
- Continuous fetal monitoring with a pulse oximeter: A case of cord compressionAmerican Journal of Obstetrics and Gynecology, 1989
- Cardiotocographs at a distance.BMJ, 1989
- MEASUREMENT OF FETAL PERIPHERAL PERFUSION WITH A PULSE OXIMETERThe Lancet, 1989
- Pulse OximetryAnesthesiology, 1989
- CONTINUOUS INTRAPARTUM MEASUREMENT OF FETAL OXYGEN SATURATIONThe Lancet, 1988
- CONTINUOUS INTRAPARTUM MEASUREMENT OF FETAL OXYGEN SATURATIONThe Lancet, 1988
- The efficacy of intrapartum electronic fetal monitoringAmerican Journal of Obstetrics and Gynecology, 1987