Right and left ventricular stroke volume measurements with velocity-encoded cine MR imaging: in vitro and in vivo validation.

Abstract
The accuracy of measurements of flow velocity determined by using cine MR phase velocity mapping--velocity-encoded cine (VEC) MR--was assessed by comparing VEC MR data with independent measurements in a flow phantom and in human subjects. Constant flow velocities generated in a phantom (range, 20-408 cm/sec) were determined correctly by VEC MR (r = .997, standard error of the estimate [SEE] = 7.9 cm/sec). Peak systolic velocities in the main pulmonary artery determined by VEC MR correlated well with the measurements obtained by using continuous-wave Doppler echocardiography (r = .91). Stroke volumes measured at the aorta by VEC MR and continuous-wave Doppler imaging also correlated well with each other (r = .80). VEC MR measurements of aortic and pulmonary flow provided left and right ventricular stroke volumes that correlated well with left ventricular stroke volumes determined by short-axis cine MR images (r = .98, SEE = 3.7 ml, and r = .95, SEE = 4.8 ml, respectively). Intra- and interobserver variabil...