Six-year experience with the St.Jude medical prosthesis: Early and late results of 540 valves in 462 patients.

Abstract
From July 1978 to June 1984, 540 St. Jude Medical valve prostheses were implanted in 462 adult patients. Overall operative mortality was 4.8% (22/462): 5.7% after aortic valve replacement (AVR) (8/140); 3.3% after mitral valve replacement (MVR) (8/244); and 7.7% after aortic and mitral valve replacement (DVR) (6/78). There were 16 late deaths among 440 operative survivors: six (38%) were valve related. The actuarial survival rate at six years was 94.5% in AVR and 93.5% in MVR. At five years it was 91.6% in DVR. The freedom from valve-related death at six years after AVR, MVR and at five years after DVR was 97.8%, 98.6% and 95.8%, respectively. The incidence of valve-related complications (expressed as %/patient-year (number of events)) was as follows: thrombo-embolism (AVR 0.7 (2), MVR 2.0(13), DVR 2.8 (5)), valve thrombosis (AVR 0.3 (1), MVR 0.2 (1), DVR 0 (0)), prosthetic valve endocarditis (AVR 0.3 (1), MVR or DVR 0 (0)). There were no cases of primary structural failure, serious anticoagulant-related hemorrhage or clinical event of hemolysis. The incidence of all valve-related complications in AVR, MVR and DVR was 1.3, 2.2 and 2.8%/pt-yr, respectively. The freedom from thrombo-embolism at six years after AVR and MVR and at five years after DVR were 91.4%, 92.4% and 88.7% respectively. Re-operation was performed in only one patient with MAR because of valve thrombosis (0.2%/pt-yr after MAR, overall 0.1%/pt-yr). These results indicate that St. Jude Medical prosthesis has provided a satisfactory performance with acceptable risks of late complications in patients who were properly anticoagulated during six-year follow-up period.