Evaluation of the Minitek System for Identification of Enterobacteriaceae
Open Access
- 1 January 1974
- journal article
- research article
- Published by American Society for Microbiology in Applied Microbiology
- Vol. 28 (4) , 668-671
- https://doi.org/10.1128/aem.28.4.668-671.1974
Abstract
Clinical isolates (869) and stock cultures (35) of Enterobacteriaceae were tested in parallel with the Minitek and conventional systems. The Minitek correctly identified 822 of 904 cultures. When a deoxyribonuclease plate was inoculated along with the Minitek, it was possible to speciate Enterobacteriaceae within 24 h. False-positive hydrogen sulfide reactions were the major fault with this system. Reactions were clear-cut and easy for technologists to read.Keywords
This publication has 0 references indexed in Scilit: