Respiration-induced Attenuation Artifact at PET/CT: Technical Considerations
- 1 March 2003
- journal article
- Published by Radiological Society of North America (RSNA) in Radiology
- Vol. 226 (3) , 906-910
- https://doi.org/10.1148/radiol.2263011732
Abstract
Combined positron emission tomographic (PET)/computed tomographic (CT) scanners allow the use of CT data for attenuation correction of PET images. Eight patients with cancer underwent PET/CT scanning. Transmission scanning was performed with conventional attenuation correction and with CT scanning during maximum inspiration and normal expiration. Image quality was visually compared and fluorine 18 activities were measured in volumes of interest in the lung and myocardium. Analysis of variance for repeated measures revealed a significant decrease (P = .0001) in measured activities between PET images corrected with CT data acquired during maximum inspiration and those corrected with the conventional attenuation correction method or with CT data acquired during normal expiration. Deep inspiration during CT can result in severe deterioration in the final PET image. © RSNA, 2003Keywords
This publication has 7 references indexed in Scilit:
- PET-CT image co-registration in the thorax: influence of respirationEuropean Journal of Nuclear Medicine and Molecular Imaging, 2002
- CT vs 68Ge attenuation correction in a combined PET/CT system: evaluation of the effect of lowering the CT tube currentEuropean Journal of Nuclear Medicine and Molecular Imaging, 2001
- Combined PET/CT Imaging in Oncology Impact on Patient ManagementClinical Positron Imaging, 2000
- FDG Imaging of Lung Nodules: A Phantom Study Comparing SPECT, Camera-based PET, and Dedicated PETRadiology, 1999
- Attenuation correction for a combined 3D PET/CT scannerMedical Physics, 1998
- A JAVA environment for medical image data analysis: Initial application for brain PET quantitationMedical Informatics, 1998
- Movements of the thoracic cage and diaphragm in respiration*The Journal of Physiology, 1954