Particulate-Associated Endosteal Osteolysis in Titanium-Base Alloy Cementless Total Hip Replacement
- 1 January 1992
- book chapter
- Published by ASTM International
Abstract
Fourteen cases of diaphyseal endosteal erosions adjacent to the femoral stem have been identified in our patient population of 308 primary cementless titanium alloy total hip replacements at minimum two-year follow-up. Three of these cases were associated with unstable prostheses that resulted in revision surgery. Eleven cases were associated with radiographically stable prostheses. one of which was revised. Capsular tissue and interfacial membrane tissue were harvested from the four cases that were revised. All the tissues were analyzed histologically. and selected tissues were analyzed with electron microprobe analysis (EMA). analytic electron microscopy (AEM). and Fourier transform infrared spectroscopy (FTIR). Standard hematoxylin and eosin sections demonstrated histiocytic infiltration of the capsular and interfacial membrane tissue. These cells were laden with fine intracellular birefringent material. EMA revealed the presence of titanium alloy particles and AEM documented titanium alloy particles as well as silicate particles. FTIR positively identified polyethylene-like particles, but this technique was not able to identify particles less than 10 μm in diameter. Femoral endosteal osteolysis appears to be mediated by histiocytes containing both metallic and polymeric wear debris.This publication has 10 references indexed in Scilit:
- Osteolysis in Alloarthroplasty of the HipClinical Orthopaedics and Related Research, 1990
- Osteolysis in Alloarthroplasty of the HipClinical Orthopaedics and Related Research, 1990
- Endosteal erosion in association with stable uncemented femoral components.Journal of Bone and Joint Surgery, 1990
- Measuring the volume fraction of bone ingrowth: A comparison of three techniquesJournal of Orthopaedic Research, 1990
- Femoral lysis after cemented hip arthroplastyThe Journal of Arthroplasty, 1988
- Localized osteolysis in stable, non-septic total hip replacement.Journal of Bone and Joint Surgery, 1986
- Localized Bone Resorption in the Femur in Mechanical Failure of Cemented Total Hip ArthroplastiesActa Orthopaedica, 1983
- ???Modes of Failure??? of Cemented Stem-type Femoral ComponentsClinical Orthopaedics and Related Research, 1979
- Reactions of the articular capsule to wear products of artificial joint prosthesesJournal of Biomedical Materials Research, 1977
- Extensive localized bone resorption in the femur following total hip replacementJournal of Bone and Joint Surgery, 1976