Dedicated PET scanners for breast imaging
- 1 December 1997
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 42 (12) , 2463-2480
- https://doi.org/10.1088/0031-9155/42/12/012
Abstract
We have used computer simulations to compare two designs for a PET scanner dedicated to breast imaging with a whole-body PET scanner. The new designs combine high spatial resolution, high sensitivity, and good energy resolution to detect small, low-contrast masses. The detectors are position sensitive NaI(Tl) scintillators. The first design is a ring scanner surrounding the breast and the second consists of two planar detectors placed on opposite sides of the breast. We have employed standard performance measures to compare the different designs: contrast, percentage standard deviation of the background, and signal-to-noise ratios of reconstructed images. The results of the simulations show that both of the proposed designs have better lesion detectability than a whole-body scanner. The results also show that contrast is higher in the ring breast system but that the noise is lower in the planar breast system. Overall, the ring system yields images with the best signal-to-noise ratios, although the planar system offers practical advantages for imaging the breast and axilla.Keywords
This publication has 14 references indexed in Scilit:
- A factorization method for the 3D X-ray transformInverse Problems, 1995
- Accelerated EM reconstruction in total-body PET: potential for improving tumour detectabilityPhysics in Medicine & Biology, 1994
- Design and performance of the HEAD PENN-PET scannerIEEE Transactions on Nuclear Science, 1994
- Figures of merit for comparing reconstruction algorithms with a volume-imaging PET scannerPhysics in Medicine & Biology, 1994
- Positron emission tomography and breast masses: Comparison with clinical, mammographic, and pathological findingsAnnals of Surgical Oncology, 1994
- Three-dimensional image reconstruction for PET by multi-slice rebinning and axial image filteringPhysics in Medicine & Biology, 1994
- Accelerated image reconstruction using ordered subsets of projection dataIEEE Transactions on Medical Imaging, 1994
- Evaluation of breast masses and axillary lymph nodes with [F-18] 2-deoxy-2-fluoro-D-glucose PET.Radiology, 1993
- Effect of increased axial field of view on the performance of a volume PET scannerIEEE Transactions on Medical Imaging, 1993
- Analytic 3D image reconstruction using all detected eventsIEEE Transactions on Nuclear Science, 1989