Cascaded linear systems based modeling techniques have been used in the past to predict important system parameters that have a direct impact on image quality. Such models are also useful in optimizing system parameters to improve image quality. In this work, detailed analysis of a solid‐state fluoroscopic imaging system intended for high‐resolution angiography is presented with the use of such a model. The imaging system analyzed through this model uses four 8×8 cm three‐side buttable interlined charge‐coupled devices (CCDs) specifically designed for high‐resolution angiography and tiled in a seamless fashion to achieve a field of view (FOV) of 16×16 cm. Larger FOVs can be achieved by tiling more CCDs in a similar manner. The system employs a CsI:Tl scintillator coupled to the CCDs by straight (nontapering) fiberoptics and can potentially be operated in 78, 156, or 234 μm pixel pitch modes. The system parameters analyzed through this model include presampling modulation transfer function, noise power spectrum, and detective quantum efficiency (DQE). The results of the simulations performed indicate that DQE(0) in excess of 0.6 is achievable, with the imager operating at 156 μm pixel pitch, 30 frames/s, and employing a 450‐μm‐thick CsI:Tl scintillator, even at a low fluoroscopic exposure rate of 1 μR/frame. Further, at a nominal fluoroscopic exposure rate of 2.5 μR/frame there was no noticeable degradation of the DQE even at the 78 μm pixel pitch mode suggesting that it is feasible to perform high‐resolution angiography hitherto unattainable in clinical practice.Keywords
This publication has 46 references indexed in Scilit:
- Parallel cascades: New ways to describe noise transfer in medical imaging systemsMedical Physics, 2001
- Mammographic imaging with a small format CCD-based digital cassette: physical characteristics of a clinical system.Medical Physics, 2000
- Light Distributions from Point, Line and Plane Sources for Photochemical Reactions and Fluorescence in Turbid Biological TissuesPhotochemistry and Photobiology, 1998
- A spatial‐frequency dependent quantum accounting diagram and detective quantum efficiency model of signal and noise propagation in cascaded imaging systemsMedical Physics, 1994
- Analysis of signal and noise propagation for several imaging mechanismsJournal of the Optical Society of America A, 1989
- Detective quantum efficiency of imaging systems with amplifying and scattering mechanismsJournal of the Optical Society of America A, 1987
- SNR and DQE analysis of broad spectrum X-ray imagingPhysics in Medicine & Biology, 1985
- Energy and angular dependence of x-ray absorption and its effect on radiographic response in screen-film systemsPhysics in Medicine & Biology, 1983
- Transfer function analysis of radiographic imaging systemsPhysics in Medicine & Biology, 1979
- Random Noise Produced by X-Ray Fluorescent Screens*Journal of the Optical Society of America, 1968